SureTemp® Thermometer Calibration Key for 690/692

Whenever the clinical accuracy of the thermometer is in question, verify the instrument calibration. It is recommended the unit be tested annually.

This method uses a fixed value to verify that the device is calibrated properly.

1. Remove any probe currently connected to the SureTemp® Plus Thermometer.

2. Connect the SureTemp® Calibration Key.

3. Wait one minute for the device to enter Monitor mode and display the temperature.

4. The temperature should read 36.3 +/- 0.1ºC or 97.3 +/- 0.2ºF.

5. Remove the SureTemp® Calibration Key.

6. Connect the original probe back to the device.